09 | Keysight | InniiVision 6000L Series Low-Prole Oscilloscopes - Data Sheet
Performance characteristics (continued)
Analog to digital conversion
Vertical resolution 8 bits
Sample rate DSO6014L: 2 GSa/sec
DSO6054L/6104L: 4 GSa/sec half channel, 2 GSa/sec each channel
Equivalent-time sample rate: 400 GSa/s (when realtime mode is turned off)
Memory depth 2 channels/4 channels
Standard 8 Mpts/4 Mpts
Time range 5 nsec/div to 50 sec/div (DSO6014L)
1 nsec/div to 50 sec/div (DSO6054L)
500 psec/div to 50 sec/div (DSO6104L)
Acquisition
Acquisition mode Normal, Peak Detect, Averaging, High Resolution
Peak detection DSO6014L: 1 nsec peak detect
DSO6054L/6104L: 250 psec peak detect
Averaging Selectable from 2,4,8,16,32,64… to 65536
High resolution mode Time base Bits of resolution
< 100 nsec/div 8
500 nsec/div 9
2 µsec/div 10
10 µsec/div 11
≥ 50 µsec/div 12
Filter Sinx/x interpolation
Trigger system
Sources DSO6xx4L: Ch 1, 2, 3, 4, line, ext and D0 - D15 for MSO enabled DSO
Modes Auto, Normal, Single
Holdoff time range ~60 ns to 10 seconds
Trigger jitter 15 psec rms
Selections Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, USB, I
2
C, SPI, Nth edge burst
Scope channel triggering
Range (internal) ±6 div from center screen
Sensitivity* < 10 mV/div: greater of 1 div or 5 mV
≥ 10mV/div: 0.6 div
Coupling AC (~10 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.